CSA C22.2 No. 60601-2-75:24
Scope
This Standard applies to ME EQUIPMENT and ME SYSTEMS that are intended to be installed or used in
accordance with CSA C22.1, Canadian Electrical Code, Part I.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this document are not covered by specific requirements in this document
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This document applies to PHOTODYNAMIC THERAPY AND PHOTODYNAMIC DIAGNOSIS EQUIPMENT
used for compensation or alleviation of disease, injury or disability.
In the case of combined equipment (e.g. equipment additionally provided with a function or an
APPLIED PART for the target area) such equipment shall also comply with any particular
standard specifying safety requirements for the additional function.
This particular standard does not apply to:
• light therapy equipment intended for use in photothermal ablation, coagulation, and
hyperthermia;
• low-level laser therapy equipment not intended for use with a PHOTOSENSITIZER;
• illumination equipment intended for use in observation, monitoring, and diagnosis, not
intended for use with a PHOTOSENSITIZER.