CSA C22.2 No. 60601-2-21:23
Scope
Replacement:
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of INFANT
RADIANT WARMERS as defined in 201.3.204, also referred to as ME EQUIPMENT.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this document are not covered by specific requirements in this document,
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This particular standard specifies the safety requirements for INFANT RADIANT WARMERS, but
alternate methods of compliance with a specific clause, by demonstrating equivalent safety,
will not be judged as non-compliant, if the MANUFACTURER has demonstrated in his RISK
MANAGEMENT FILE that the RISK presented by the HAZARD has been found to be of an
acceptable level when weighed against the benefit of treatment from the device.
This particular standard does not apply to:
• devices supplying heat via BLANKETS, PADS or MATTRESSES in medical use; for information,
see IEC 60601-2-35;
• INFANT INCUBATORS; for information, see IEC 60601-2-19;
• INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20;
• INFANT PHOTOTHERAPY EQUIPMENT, for information, see IEC 60601-2-50.
SKIN TEMPERATURE SENSORS which are applied to operate a BABY CONTROLLED RADIANT WARMER
including the displayed value are not considered to be a CLINICAL THERMOMETER in the sense
of the particular standard ISO 80601-2-56.