CSA C22.2 No. 61010-2-040:21
1 Scope and object
This clause of Part 1 is applicable except as follows:
1.1.1 Equipment included in scope
Replacement:
Replace the existing text with the following:
This part of IEC 61010 specifies safety requirements for electrical equipment intended for
sterilization, washing, and disinfection of medical materials in the medical, veterinary,
pharmaceutical and laboratory fields, when used under the environmental conditions of 1.4.
Examples of such equipment include the following:
a) STERILIZERS and disinfectors using steam and/or hot water as the sterilant;
b) STERILIZERS and disinfectors using toxic gas, toxic aerosol or toxic vapour as the sterilant;
c) STERILIZERS and disinfectors using hot air or hot inert gas as the sterilant; and
d) WASHER-DISINFECTORS.
1.1.2 Equipment excluded from scope
Addition:
Add the following note to item f):
NOTE IEC 60601-1:2005, 3.63, defines "medical electrical equipment" as follows (notes to entry are omitted):
Electrical equipment, having an applied part or transferring energy to or from the patient or detecting such energy
transfer to or from the patient and which is:
a) provided with not more than one connection to a particular supply MAINS; and
b) intended by its manufacturer to be used:
1) in the diagnosis, treatment, or monitoring of a patient; or
2) for compensation or alleviation of disease, injury or disability.
Addition:
Add the following new second paragraph after the lettered list:
This document does not apply to the following types of equipment:
aa) equipment for use in hazardous atmospheres (see IEC 60079); however this document
does apply to an atmosphere created inside equipment by a flammable sterilizing
agent (see 13.2.101 and 13.2.102);
bb) laboratory equipment for the heating of materials for purposes other than sterilization
or disinfection (see IEC 61010-2-010);
cc) laundry equipment (see IEC 60335-2-4, IEC 60335-2-7, IEC 60335-2-11, and
ISO 10472 (all parts)), unless designed for disinfecting medical materials;
dd) dishwashers (see IEC 60335-2-5 and IEC 60335-2-58).
1.2.1 Aspects included in scope
Replacement:
Replace item g) with the following new text:
g) liberated gases (including the non-intentional escape of toxic gas), pathogenic
substances, explosion and implosion (see Clause 13).
1.2.2 Aspects excluded from scope
Addition:
Add the following two new items:
aa) special requirements for protection against chemical and high-risk micro-biological
HAZARDS associated with the LOAD;
bb) general requirements for the design of calorifiers, shell boilers and PRESSURE VESSELS.
NOTE National and other regulations or codes apply for the safety of calorifiers, shell boilers and
PRESSURE VESSELS (see 14.101).
This Standard applies to the above-noted equipment that is intended to be installed or used in
accordance with CSA C22.1, Canadian Electrical Code, Part I.
General requirements applicable to the above-noted equipment are provided in CSA C22.2 No. 0.
OEN:
CSA
Langue:
English
Code(s) de l'ICS:
19.080;
71.040.10
Statut:
Norme
Date de Publication:
2021-02-28
Numéro Standard:
CSA C22.2 No. 61010-2-040:21