CSA C22.2 No. 80601-2-74:19
Scope
Replacement:
This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of a HUMIDIFIER, also
hereafter referred to as ME EQUIPMENT, in combination with its ACCESSORIES, the combination also
hereafter referred to as ME SYSTEM.
This document is also applicable to those ACCESSORIES intended by their MANUFACTURER to be
connected to a HUMIDIFIER where the characteristics of those ACCESSORIES can affect the BASIC
SAFETY or ESSENTIAL PERFORMANCE of the HUMIDIFIER.
EXAMPLE 1 Heated BREATHING TUBES (heated-wire BREATHING TUBES) or ME EQUIPMENT intended to
control these heated BREATHING TUBES (heated BREATHING TUBE controllers).
NOTE 1 Heated BREATHING TUBES and their controllers are ME EQUIPMENT and are subject to the
requirements of IEC 60601-1.
NOTE 2 ISO 5367 specifies other safety and performance requirements for BREATHING TUBES.
This document includes requirements for the different medical uses of humidification, such as
invasive ventilation, non-invasive ventilation, nasal high-flow therapy, and obstructive sleep
apnoea therapy, as well as humidification therapy for tracheostomy PATIENTS.
NOTE 3 A HUMIDIFIER can be integrated into other equipment. When this is the case, the requirements of
the other equipment also apply to the HUMIDIFIER.
EXAMPLE 2 Heated HUMIDIFIER incorporated into a critical care ventilator where ISO 80601-2-12[12]
also applies.
EXAMPLE 3 Heated HUMIDIFIER incorporated into a homecare ventilator for dependent PATIENTS where
ISO 80601-2-72[14] also applies.
EXAMPLE 4 Heated HUMIDIFIER incorporated into sleep apnoea therapy equipment where ISO 80601-2-
70[13] also applies.
This document also includes requirements for an ACTIVE HME (HEAT AND MOISTURE EXCHANGER),
ME EQUIPMENT which actively adds heat and moisture to increase the humidity level of the gas
delivered from the HME to the PATIENT. This document is not applicable to a passive HME, which
returns a portion of the expired moisture and heat of the PATIENT to the respiratory tract during
inspiration without adding heat or moisture.
NOTE 4 ISO 9360-1[5] and ISO 9360-2[6] specify the safety and performance requirements for a passive
HME.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within
the scope of this document are not covered by specific requirements in this document except in
IEC 60601-1:2005+AMD1:2012, 7.2.13 and 8.4.1.
NOTE 5 Additional information can be found in IEC 60601-1:2005+AMD1:2012, 4.2.
This document does not specify the requirements for cold pass-over or cold bubble-through
humidification devices, the requirements for which are given in ISO 20789:—.[8]
This document is not applicable to equipment commonly referred to as “room humidifiers” or
humidifiers used in heating, ventilation and air conditioning systems, or HUMIDIFIERS
incorporated into infant incubators.
This document is not applicable to nebulizers used for the delivery of drugs to PATIENTS.
NOTE 6 ISO 27427[10] specifies the safety and performance requirements for nebulizers.
This document is a particular standard in the IEC 60601-1 and the ISO/IEC 80601 series.
This Standard applies to ME EQUIPMENT and ME SYSTEMS that are intended to be installed or used in
accordance with CSA C22.1, Canadian Electrical Code, Part I.
OEN:
CSA
Langue:
English
Code(s) de l'ICS:
11.040.10
Statut:
Annulée
Date de Publication:
2019-01-31
Numéro Standard:
CSA C22.2 No. 80601-2-74:19