CSA C22.2 No. 61010-2-033:20
1 Scope and object
This clause of Part 1 is applicable except as follows:
1.1.1 Equipment included in scope
Replace the existing text with the following:
This part of IEC 61010 specifies safety requirements for hand-held multimeters for domestic
and professional use, capable of measuring MAINS.
Hand-held multimeters are multi-range multifunction measuring instruments intended to
measure voltage and other electrical quantities such as resistance or current. Their primary
purpose is to measure voltage on a live MAINS. They are suitable to be supported by one hand
during NORMAL USE.
1.1.2 Equipment excluded from scope
Add the following new item to the list and the following paragraph:
aa) IEC 61557-1 to IEC 61557-12, Electrical safety in low voltage distribution systems up to
1 000 V a.c. and 1 500 V d.c. – Equipment for testing, measuring or monitoring of
protective measures
HAND-HELD EQUIPMENT such as oscilloscopes, wattmeters, process control multimeters not
RATED for measuring voltage on a live MAINS, clamp multimeters and communications test sets
are not within the scope of this document.
1.2.1 Aspects included in scope
Add the following two new paragraphs at the end of the subclause:
Requirements for protection against HAZARDS resulting from NORMAL USE and REASONABLY
FORESEEABLE MISUSE of measuring circuits are given in Clause 101.
Requirements for reliance on the displayed value are given in Clause 102.
This Standard applies to the above-noted equipment that is intended to be installed or used in
accordance with CSA C22.1, Canadian Electrical Code, Part I.
General requirements applicable to the above-noted equipment are provided in CSA C22.2 No. 0.
OEN:
CSA
Langue:
English
Code(s) de l'ICS:
19.080;
71.040.10
Statut:
Norme
Date de Publication:
2020-07-31
Numéro Standard: