CAN/CSA-C108.6-M91 (R2003)
This National Standard of Canada is equivalent to International Standard IEC CISPR 11:1990.
1.1
The limits and methods of measurement laid down in this publication apply to
industrial, scientific and medical (ISM) equipment as defined in Clause 2, and
to spark erosion equipment.
1.2
The assessment of conformity of equipment tested on a test site shall be in
accordance with the specifications of Clause 6. For equipment in series
production, there shall be 80% confidence that at least 80% of manufactured
items comply with the limits given. The statistical assessment procedure is
specified in Sub-clause 6.3. For small scale production the application of the
statistical assessment may not always be economical. In such cases the
assessment procedure contained in Sub-clause 6.2 may be applied. Measurement
results obtained for an equipment measured in its place of use and not on a
test site shall relate to that installation only, and shall not be considered
representative of any other installation and so shall not be used for the
purpose of a statistical assessment.
Note. - The limits have been determined on a probabilistic basis taking into
account the likelihood of interference. In cases of interference,
additional provisions may be required.
1.3
Procedures are given for the measurement of radio frequency disturbances and
limits are laid down within the frequency range 9 kHz to 400 GHz.
1.4
Requirements for lighting apparatus are contained in CISPR 15.
OEN:
CSA
Langue:
English
Code(s) de l'ICS:
17.220.20;
33.100
Statut:
Annulée
Date de Publication:
1991-12-30
Numéro Standard: