CSA C22.2 No. 61010-2-010:19
1 Scope and object
This clause of Part 1 is applicable except as follows:
1.1.1 Equipment included in scope
Replacement:
This part of IEC 61010 specifies safety requirements for electrically powered laboratory
equipment for the heating of materials, where the heating of materials is one of the functions
of the equipment.
NOTE If all or part of the equipment falls within the scope of one or more other part 2 standards of IEC 61010 as
well as within the scope of this standard, it will also need to meet the requirements of those other part 2 standards.
In particular, if equipmen is intended to be used for IVD purposes, it will need to meet the requirements of
IEC 61010-2-101.
1.1.2 Equipment excluded from scope
Addition after item j):
aa) equipment for the heating and ventilation of laboratories;
bb) sterilizing equipment;
cc) heating and/or cooling equipment which the OPERATOR is intended to enter, and which is
large enough for the OPERATOR to remain inside with the door or doors closed.
This Standard applies to the above-noted equipment that is intended to be installed or used in
accordance with CSA C22.1, Canadian Electrical Code, Part I.
General requirements applicable to this Standard are provided in CAN/CSA-C22.2 No. 0.
OEN:
CSA
Langue:
English
Code(s) de l'ICS:
19.080;
71.040.20
Statut:
Norme
Date de Publication:
2019-10-31
Numéro Standard: