CAN/CSA-ISO 11712:14
1 Scope
1.1 This International Standard provides the essential requirements for the design of supralaryngeal
airways and connectors. These devices are intended to open and seal the supralaryngeal area to provide an
unobstructed airway in patients during spontaneous, assisted or controlled ventilation.
1.2 This International Standard specifies the dimensions, basic properties and method of size designation
of the available types of supralaryngeal airways. Airways devised for specialized applications are not
specifically covered, although most may be classified by the sizing and dimensions (or other characteristics)
required by this International Standard.
1.3 The following devices are outside the scope of this International Standard: nasal and oropharyngeal
airways, anesthetic masks, oro- and naso-tracheal tubes, cricothyrotomy devices, dental appliances, tracheal
stents, tracheal tubes, ventilating laryngoscopes, CPAP devices, esophageal obturators, bougies and devices
that require surgical placement.
1.4 This International Standard requires dimensional disclosure so the operator will know which auxiliary
instruments, such as tracheal tubes and bronchoscopes will be size-compatible.
1.5 Flammability of airways, for example if used with certain flammable anesthetics, electrosurgical units or
lasers, is a well-recognized hazard that is outside the scope of this International Standard. See E.1.7.