CAN/CSA-Z314.8-00 (R2007)
1. Scope
1.1
This Standard describes essential elements and methods in the safe handling,
transportation, and biological decontamination of contaminated reusable medical
devices (see Figure 1).
Note: The biological decontamination process includes thorough cleaning and,
whenever necessary for personnel or patient safety, appropriate application of
a microbicidal process (disinfection or sterilization).
1.2
This Standard includes requirements for
(a) qualification, education, and training of personnel;
(b) design of decontamination areas;
(c) protective attire;
(d) immediate handling of contaminated medical devices at the point of use;
(e) transport of contaminated medical devices;
(f) emergency preparedness (spills); and
(g) decontamination policies and procedures (in general, as related to reusable
diagnostic and therapeutic medical devices, and specifically, as related to
reusable respiratory/anaesthesia devices, flexible fibre optic endoscopes, and
electronic patient-applied devices).
Note: It is recognized that more detailed procedures may be required for
specific medical devices or applications.
1.3
This Standard does not apply to
(a) the terminal sterilization of reusable medical devices;
Note: Refer to CSA Standards CAN/CSA-Z314.2, CAN/CSA-Z314.3, and ISO Draft
Standard ISO/DIS 14937.
(b) single-use/disposable medical devices;
Note: Information concerning safety, technology, cost/benefit, and legal issues
involving the reuse of such devices is found in such publications as the
Canadian Healthcare Association's The Reuse of Single-Use Medical Devices:
Guidelines for Healthcare Facilities and ECRI's Reuse of Medical Devices:
Making Informed Decisions.
(c) medical devices that have been used with patients who are known or
suspected to have Creutzfeldt-Jakob Disease or prion-related diseases; and
Note: Refer to the following Health Canada publication: Infection Control
Guidelines for Healthcare Workers for Creutzfeldt-Jacob Disease in Canada (in
progress).
(d) the reprocessing and reuse of haemodialyzers.
Note: Refer to CSA Standard CAN/CSA-Z364.1.3.
1.4
In this Standard, "shall" indicates a mandatory requirement; "should" indicates
a recommendation or that which is advised but not mandatory; and "may"
indicates an advisory or optional statement. Notes accompanying clauses do not
include mandatory or alternative requirements; the purpose of a note
accompanying a clause is to separate from the text explanatory or informative
material that is not properly a part of the Standard. Notes to figures and
tables are considered part of the figure or table and may be written as
mandatory requirements; legends to figures are also considered part of the
requirements of the figure.