CAN/CSA-ISO 7767-98 (R2003)
The international community has decided to make this standard available only in the English language version. It has been adopted as a National Standard of Canada on this basis.
This National Standard of Canada is equivalent to International Standard ISO 7767:1997.
1.1 Scope
ISO 7767 is one of a series of International Standards based on IEC 601-1 (the
"General Standard"). This type of International Standard is referred to as a
"Particular Standard". As stated in 1.3 of IEC 601-1:1988, the requirements of
this International Standard take precedence over those of IEC 601-1.
The scope given in clause 1 of IEC 601-1:1988 applies except that 1.1 shall be
replaced by the following:
This International Standard specifies safety requirements for oxygen monitors,
as defined in clause 1.3.14, intended for use in determining the oxygen level
in gas mixtures. Both diverting and non-diverting oxygen monitors are covered.
The field of application includes, but is not limited to:
a) anaesthetic machines and breathing systems;
b) ventilators;
c) infant incubators;
d) oxygen concentrators.
Devices that do not "measure and indicate" are not intended to be covered by
this International Standard. For instance, a device that has no function other
than to signal an alarm at a specific oxygen level would not be considered to
be an oxygen monitor for use in direct patient monitoring applications.
The phrase "in a gaseous mixture" implies that devices that measure or monitor
oxygen in a liquid phase (for example, blood gas analyzer or indwelling
catheters) are not covered by this International Standard.
Oxygen monitors intended for use in laboratory research applications are
outside the scope of this International Standard.