Appareils électromédicaux - Partie 2-66: Exigences particulières pour la sécurité de base et les performances essentielles des aides auditives et des systèmes d'aides auditives

Numéro de référence:
CSA C22.2 NO. 60601-2-66
Catégorie de norme:
Norme nationale du Canada - Adoption d'une Norme internationale
Type d’activité d’élaboration de normes:
Nouvelle version
Code ICS:
11.180.15
17.140.50
Statut:
En cours d'élaboration
Date de début de la période de commentaires OEN:
Date de fin de la période de commentaires des OEN:
Affiché le:

Porté:

Champ d’application

This part of IEC 60601 applies to the basic safety of hearing aids and hearing aid systems, hereafter also referred to as ME EQUIPMENT or ME SYSTEM.

 

If a clause or subclause is specifically intended to be applicable to hearing aids only, or to hearing aid systems only, the title and content of that clause or subclause will say so. if that is not the case, the clause or subclause applies both to hearing aids and to hearing aid systems, as relevant.

 

Hazards inherent in the intended physiological function of hearing aids or hearing aid systems within the scope of this document are not covered by specific requirements in this document except in 201.7.9.2 and 201.9.6.

 

NOTE See also 4.2 of the general standard.

 

Accessories to hearing aids in the home healthcare environment (e.g. remote control units, audio streamers, battery chargers, power supplies) can be tested according to the applicable standard, IEC 60065, IEC 60950-1, IEC 62368-1 or other applicable IEC safety standards. Alternatively, the general standard may be applied. Hearing aids do not have a mains part intended for connection to ac supply mains. The connection to the supply mains of a hearing aid system is covered by power supply, charger or other types of accessories.

 

Accessories with functional connection to a hearing aid may form a hearing aid system. Hearing aid related accessories that are not physically connected to the hearing aid during normal use are not considered to be applied part, because they do not directly contribute to the intended use of the hearing aid.

 

Wireless programming interfaces are covered by the applicable standard IEC 60065, IEC 60950-1, IEC 62368-1 or other applicable IEC safety standards. Alternatively, the general standard may be applied.

 

Programming interfaces with wired connection to the hearing aid are covered by the general standard.

 

NOTE Detachable parts of hearing aids, even if supplied separately (e.g. ear hooks, domes, wax guards etc.), are not considered as accessories, but as component parts.

 

This document does not apply to:

•             Cochlear implants or other implanted hearing aids;

•             Bone conduction hearing aids;

•             Educational hearing aids (i.e. group hearing aids, auditory trainers etc.);

•             The application of a hearing aid for the measurement of hearing levels; IEC 60645-1 applies; – fix installed audio-frequency induction-loop systems or their component parts, as described in IEC 60118-4 and IEC 62489-1;

•             The sound generating function of a tinnitus masker.

 

This document does not address applicable testing for intentional RF radiation of wireless equipment (e.g. maximum radiated output power, modulation bandwidth, etc.).

Raison d’être du projet

Raison d’être du projet
To maintain alignment with international requirements on the particular requirements for the basic safety and essential performance of external cardiac pacemakers with internal power source.

Note : L’information ci-dessus a été recueillie et est diffusée par le Conseil canadien des normes (CCN) pour les besoins de son système de notification centralisé et transparent pour l’élaboration de nouvelles normes. Le système permet aux organismes d’élaboration de normes (OEN) accrédités par le CCN et aux membres du public d’être informés des nouveaux travaux d’élaboration de normes au Canada. Il donne aussi aux OEN accrédités la possibilité de repérer et de résoudre les cas de doubles emplois éventuels dans les normes et les travaux de normalisation.

Les OEN sont eux-mêmes responsables du contenu et de l’exactitude de l’information présentée ici. Cette information n’existe que dans la langue dans laquelle elle a été fournie au CCN.